Product

Semi-micro cuvette, 3 ml, (HxW): 45 x 12 mm, PS, transparent, optical sides: 2

Order number: 67.746
Semi-micro cuvette, 3 ml, (HxW): 45 x 12 mm, PS, transparent, optical sides: 2

Product description

Order number 67.746
Product description Semi-micro cuvette, max. working volume: 3 ml, material: PS, transparent, (HxW): 45 x 12 mm, optical sides: 2, path length: 10 mm, 500 piece(s)/bag

Product characteristics

Volume of work 1 ml - 3 ml
Number of optical sides 2
Path length 10 mm
Wavelength in the cuvette area 330 nm

Size

Max. volume 3 ml
Width of product 12 mm
Height of product 45 mm

Material & colours

Product material Polystyrene (PS)
Colour of product transparent

Purity & certification

Product category no medical device | IVD
Batched no

Packaging

Minimum order qty. 2000
Type of smallest subpackaging bag
Piece(s) / inner box 500
Piece (s) / outer case 2000
Piece(s) / pallet 96000
Depth of case 388 mm
Width of case 378 mm
Height of case 215 mm
Case volume 0.0315 cbm
Weight of product 0.0023 kg
Weight of case 5.02 kg
EAN of inner box 4038917083928
EAN case 4038917002806
Product specification

General

  • +

    Can the 80.625 swab be autoclaved?

    Our 80.625 swab cannot be autoclaved. The tube in the swab is made out of the autoclavable plastic polypropylene, but both the cap (polyethylene) and the swab stem (polystyrene) are not autoclavable. This is due to the low temperature resistance of the PE and PS plastic, which have a maximum long-term use temperature of 60–80°C. Autoclaving is carried out at temperatures of at least 121°C and it is not possible to rule out an impairment of the product as a result of this.

  • +

    How is the centre height (Z) of cuvettes defined?

    The distance between the base of the cuvette and the centre of the photometer light beam is defined as the centre height (Z). The centre height can vary depending on the manufacturer and the type of photometer, which is why Sarstedt offers UV cuvettes in two different centre heights (Z = 8 mm and Z = 15 mm). If customers do not know the exact centre height of the photometer, this information can be requested from the device manufacturer.

  • +

    Are there cuvettes which are suitable for fluorescence measurements?

    Fluorescence measurements are carried out at a 90° angle to the direction of the photometer beam and can therefore only be used in quadruple optical cuvettes. We offer a quadruple optical cuvette made of polystyrene (67.754) or PMMA (67.755) especially for this application.

  • +

    In which wavelength range can Sarstedt cuvettes be used?

    Thanks to their optical properties, our polystyrene cuvettes can be used from the visual range to the UV range to 330 nanometres (nm). In contrast to this, acrylic (PMMA) cuvettes are characterised by a high level of transparency in the UV range and can therefore be used to approximately 300 nm. The plastic material of the UV cuvette enables measurements in the UV range from a wavelength of 220 nm.

Go to top