Order number | 95.902 |
---|---|
Product description | Post transport packaging, application e.g.: external packaging according to P 650/ ADR, flat design, (LxWxH): 220 x 170 x 40 mm, for diagnostic specimens, holds 1 to 4 transport containers/bottles up to a diameter of 30 mm, enclosed foam foil pouch, 50 piece(s)/case |
UN identification | yes |
Label/ Print | with print |
---|---|
Colour of print/label | black |
Design | flat design |
Example of use | external packaging according to P 650/ ADR |
Suitable for | diagnostic specimens |
Width of product | 170 mm |
---|---|
Height of product | 40 mm |
Length of product | 220 mm |
Product material | cardboard |
---|---|
Colour of product | brown |
Satisfies the requirement | P 650, ADR, IATA |
---|---|
Product category | no medical device | IVD |
Batched | yes |
Minimum order qty. | 50 |
---|---|
Type of smallest subpackaging | case |
Piece(s) / inner box | 50 |
Piece (s) / outer case | 50 |
Piece(s) / pallet | 2400 |
Depth of case | 488 mm |
Width of case | 401 mm |
Height of case | 145 mm |
Case volume | 0.0284 cbm |
Weight of product | 0.0679 kg |
Weight of case | 3.86 kg |
EAN of inner box | 4038917084574 |
EAN case | 4038917069854 |
The foam attached inside the transport case cannot be replaced.
In accordance with the packaging instructions P650, packaging for UN3373 must consist of at least three components:
- a primary tube
- a secondary tube
- outer packaging.
It is necessary to ensure that the two tubes are impervious to liquids (if liquid substances are being shipped) and one of the two has an internal pressure of 95 kPa (0.95 bar) with no loss of filling material.
An absorbent material which can soak up the entire filling quantity must be inserted between the two tubes.
At least one of the surfaces of the outer packaging must have minimum dimensions of 100 x 100 mm.
Wipe disinfection is recommended as a method of cleaning as this method rules out the risk of the product being subjected to mechanical or thermal stress.
In accordance with the regulations of Deutsche Post AG and the packaging directive P650 IATA/ADR, the mailing containers are a valid shipping system.
The mailing containers can be used for all Sarstedt mailing containers, and their integrated locking mechanism and the attached foam bag offer optimal fixation and padding of the secondary tube.
This is an exemption under category B (infectious substances allocated UN3373), but only applies if there is a professional assessment (e.g. from a doctor) stating that the patient samples to be shipped have no or minimal likelihood of containing a pathogen.
This assessment should be made on the basis of the known patient history, symptoms and individual background of the patient.
When being shipped, the samples must be packaged in accordance with P650. This exemption is not subject to the regulations of ADR and can therefore be transported with a courier driver with the label "EXEMPT HUMAN SPECIMEN" or "EXEMPT ANIMAL SPECIMEN".
The transport cases have a confirmation from BAM (Bundesanstalt für Materialforschung und -prüfung, Federal Institute for Materials Research and Testing) that they can be shipped in accordance with P650 ADR. Certification of this can be requested from customer services.
We recommend that cool transport containers are disposed of after five years. If you wish to dispose of the container, please ask for the safety data sheet.
You can use this data sheet to contact your regional disposal company, which will be able to give you information on how to dispose of the container in the most environmentally friendly way possible.